Origin |
Korea |
Model Number |
2120V Cr |
Brand Name |
Optizen |
Product Description
OPTIZEN 2120V Cr
Hexachrometer OPTIZEN 2120V Cr is to analyze the hexavalent chrome which is regulated
by a lot of treaties such as RoHS, WEEE, ELV(EU), Resource sound utilize promoting law(Japan),
Electronics pollution prevention law(China) and so on.
Product Feature
Customers who make the electronics, motors(and so on) and the parts of them using
hexavalent chrome for plating and etc must check the emission volumes of hexavalent chrome.
Hexachrometer OPTIZEN 2120V Cr is launched to make your hexavalent chrome analysis
more comfortable and easy.
Product Specification / Models
- Spectral Band pass width : < 1nm
- Wavelength Accuracy : < ± 0.5nm
- Wavelength Reproducibility : < ± 0.1nm
- Photometric Range : -0.1 to 3.0A, 0 to 125%
- Photometric Accuracy : < 1% at 1A
- Photometric Reproducibility : ± 0.001A
- Stray Light : 0.05% T at 220 & 340 nm
- Stability : ±0.002 A/hr
- Light Source : Tungsten and Halogen lamp
- Display Mode Backlight, Graphical, Liquid Crystal Display with Contrast Control
- Sample Holder : Automatic Rotary type 8-position Multi-Cell Holder
- Interface Ports : RS-232C Serial and Centronics ports
- Wavelength Slew Rate : 5000nm/min
- Standard Capability : Absorbance / % transmittance, concentration and file manager
- Monochromator : Modified Czerny-turner type With 1200 lines/mm Blazed grating
- Noise Level : < 0.001A
- Drift : 0.002A/hr at 340 nm
|