Mecasys Co., Ltd.
 
Spectrophotometer

Opizen 2120UV Plus Spectrophotometer

Inquire now

Opizen 2120UV Plus Spectrophotometer

Click to enlarge image

 Origin Korea
 Model Number 2120UV plus
 Brand Name Optizen

Description

 

Optizen-2120UV plus

 

  Basic experiment! Special research! Various analyses with Split Beam UV/VIS Spectrophotometer!

 

 

OPTIZEN 2120UV plus self-splits into even power of gain of photodiode preamplifier into optimal ranges with 1nm bandwidth.

Built-in calculating programs are always updated and timely new to stay as a pioneer. A specialized noise reduction gives our OPTIZEN 2120UV plus maximum stability for voltage and signal processing.

Auto blank measuring system using 8cells auto-rotated holder provides more stable data compared with a traditional split beam spectrophotometer.

OPTIZEN 2120UV plus has a user-friendly simple interface that guides you through computations with easy-to-see touch button.

 

 

FEATURES

 

- Very optimal ranges and stability in measuring

 

- 1nm Bandwidth

 

- User-friendly-software with scanning and time course capabilities

 

- Auto blank check

 

- S.T.M. system (Select Test Measure)

 

- Practicable use for kits from different manufacturers

 

- Large LCD to display graphics and calculations

 

- 8 cells auto rotating system holder : Standard

 

- More than 100 program recordable, pre-programmable, customizable

 

- Versatile data processing

 

 

Hardware Specification

 

- Photometric System : Split Beam Optics

- Wavelength Range : 190nm to 1100nm

- Spectral Band pass width : < 1nm

- Wavelength Accuracy : < ± 0.5nm

- Wavelength Reproducibility : < ± 0.1nm

- Monochromator : Modified Czerny-turner type with 1200lines/mm blazed grating

- Stray Light : 0.05% T at 220nm & 340nm

- Photometric Range : -0.1 to 3.0A, o to 125%

- Photometric Accuracy : < 1% at 1A

- Photometric Reproducibility : ± 0.001A

- Noise Level : < 0.001A

- Drift : 0.002A/hr at 340nm

- Stability : ±0.002 A/hr

- Light Source : Tungsten-Halogen lamp and Deuterium lamp

- Display Mode : Backlight, Graphical, Liquid Crystal Display with Contrast Control

 




Inquire now